Clinical Chemistry Analyzer

Clinical Chemistry Analyzer Model CA-400

Mid-range throughput (400 tests/hour, 560 tests/hour with ISE).
Main instrument for midsize workload laboratory or backup unit for high workload laboratory.

Desktop analyzer with throughput of 400 tests/hour. Compact size and excellent performance with high accuracy and functionality. Performs regular measurement, emergency samples, and specific tests. Also appropriate as a backup analyzer for large labs.

Features

  • The answer to your laboratory needs!

      • Highly-Stable Refrigerated Reagent Tray
        Removable carousel for easy loading, 60 cooled positions to enhance reagent stability.
      • Automatic Washing System
        Effective cleaning method minimizes carry over and contamination.
      • Low Reaction Volume Cuvettes
        Efficient use of reagent with minumum reaction volume of 150㎕.
      • High Precision Sample Pipetting System
        Low sample volume of just 2.0㎕(minimum).
        Capable of processing various sample types.
        Liquid level and clash detection.
        High precision adjustable sample clot detection (Optional).
      • Multi-Functional Sample Tray
        Capable of holding primary tubes, sample cups and pediatric cups.
      • STAT Sampling
        Dedicated loading port enables quick process of emergency samples.
      • Integrated ISE Module (optional)
        Measurement of Na, K and Cl.
        Throughput of 240 tests/hour.
  • User Friendly Interface

    Intuitive GUI provides fast and simple operation for any level of operator.

    • High operability with easy-to-understand layout
    • Effective use of informative graphics and charts
    • Comprehensive assessment of analyzer condition
    • Run-Monitor
      Visually shows sample status
    • Quallity Control(QC)
      Displays QC results in various formats.
    • Callibration
      Manages calibration settings and calibration curves on one screen.
    • Results
      Results can be displayed, printed out, or saved as electronic files.

Specifications

GENERAL

Type
Desktop fully automated clinical chemistry analyzer
Usage
General chemistry as photometric assay
Immunology as photometric assay (Latex reagents available)
Assay type
End point, 2 point end, rate, 2 point rate and electrolytes
Throughput
400 tests per hour (Up to 560 tests per hour with optional ISE module)
Incubation time
10 minutes after applying sample for one-reagent assay
5 minutes after applying second reagent for two-reagent assay
Sample type
Serum, plasma and urine
Sample input system
Type:
Removable tray with sample tube holder

Capacity:
Up to 70 samples (handling of STAT available)

QC sample and Calibrator:
20 positions on designated inner tray
Sampling pipette
Type:
Micro pipette with liquid level detector and clot detector

Rinsing:
Inside and outside with purified water
Sampling pump
Type:
Micro syringe

Volume:
2.0 to 35.0 μl (increment by 0.1 μl)
Reagent system
Type:
Removable tray with reagent bottle holder on a turntable

Capacity:
60 reagent positions on a tray (30 each for 70 ml and 20 ml bottles)

Inventory:
Calculation of remaining reagent volume available

Cooling:
Cooling with peltier element (8 to 15 degrees Celsius)
Reagent pipette
Numbers of pipettors:
2

Type:
Micro pipette with liquid level detector

Rinsing:
Inside and outside with purified water

Other function:
Detection of remaining reagent volume available
Reagent pump
Type:
Micro syringe

R1 Volume:
20 to 350 μl (increment by 1 μl)

R2 Volume:
20 to 350 μl (increment by 1 μl)
Cuvette system
Number of cuvettes:
90 cuvettes on reaction line

Type:
Semi-disposable cuvette with onboard washing system

Material:
Glass

Volume for measurement:
Minimum 150 μl

Maximum volume:
450 μl in total
Reaction system
Type:
Direct heating system

Temperature:
37.0 degrees Celsius
Detector
Method:
Direct measurement of absorbance in cuvette
Bi-chromatic or Mono-chromatic

Wavelength:
340 nm, 380 nm, 415 nm, 450 nm, 510 nm, 546 nm, 570 nm, 600 nm, 660 nm, 700 nm, 750 nm, 800 nm (12 wavelengths)

Light source:
Halogen tungsten lamp
Stirring system
Type:
Stick type rotating stirrer
Dimensions
970 mm (W) x 690 mm (D) x 580 mm (H) (Analyzer only)
Option
ISE module, external tank sensor, clot detection sensor

Reagents for clinical chemistry

Reagents designed for automatic chemistry analyzer and optimised chemistry parameters offer the precise and reliable data to the laboratory.
 
  • Reagents
    Clinical chemistry reagents, Immunoturbidimetric reagents
  • Calibrators
    Clinical chemistry calibrator, Immunoturbidimetric calibrator
  • Quality control samples
    Clinical chemistry quality control sample, Immunoturbidimetric quality control sample
  • Utility reagent
    Analyzer consumables

More details

Furuno Clinical Chemistry Analyzer

  • We provide extensive support to our customers based on our extensive knowwledge of both reagents and instruments.

    MODEL Features
    CA-270 Compact desk-top instrument with 270 test/hour (maximum 450 tests/hour with ISE).
    > 8 key points in the CA-270 (document download)
    CA-400 Mid-range desk-top instrument with 400 test/hour (maximum 560 tests/hour with ISE).
    CA-800 800 tests/hour throughput capability (maximum 1200 tests/hour with ISE).

Registered CA series in UNIDO's STePP

Furuno's CA series has been registered in STePP by UNIDO (United Nations Industrial Development Organization), Tokyo Office.

The STePP (Sustainable Technology Promotion Platform) is an online database to introduce outstanding Japanese technologies for sustainable industrial development in developing and emerging countries. In this STePP, Furuno's CA-series was evaluated as a product that contributes to achieving the goals of SDGs 3 [Good health and well-being] and 9 [Industry, innovation and infrastructure].

http://www.unido.or.jp/en/technology_db/9556/

Source: UNIDO Investment and Technology Promotion OfficeTokyo website

Documentation

* Specifications subject to change without notice.