Clinical Chemistry Analyzer

Clinical Chemistry Analyzer Model CA-90 Discontinued Products

Compact size and direct operation by built-in tough panel.

Compact size analyzer for general chemistry/TDM/electrolyte. Use of disposable cuvettes enables to reduce purified water consumption. Runs operation without installing any water purification system. Disposable plastic cuvettes with automatic cuvette loading and disposal system. Easy operation by built-in touch panel without connecting to external operational PC.

Features

  • 90 tests/hour (max. 156 tests/hour with ISE)

  • Reduces purified water consumption by use of disposable cuvettes.

  • Easy operation by built-in touch panel LCD

  • Compact size with user-friendly, graphical User Interface software

  • Provides accurate TDM/DOA measurement results.

  • Minimum volume required for measurement: 120 ul

Specifications

GENERAL

Assay type
End point, 2 point end, Rate, 2 point rate and electrolytes
Throughput
90 tests/hour (max. 156 tests/hour with ISE)
Sample volume
2 to 35ul (increment by 0.1ul)
Reagent volume
20 to 350ul (increment by 1ul)
Reaction volume
120 to 450ul
Samples on board
10 samples + 1 emergency sample (max.)
Sample container
Sample tubes and sample cups
Reagents on board
20 reagent bottles
Reagent container
20ml and 50ml bottles
Reaction method
Random access discrete
Reaction container
Disposable resin cuvettes
Reaction time
10 minutes (Reagent 1: 5 min. + Reagent 2: 5 min.)
Temperature control
Dry bath type
Incubator temperature
37 degrees C ± 0.3 degrees C
Stirring speed
5 speed levels available
Wavelengths
8 wavelengths (340 to 800nm)
Photometric assay
Filters
External output
RS-232C
Weight
Approx. 65kg
Dimensions
620(W) ×551(D)×577(H) mm